S IMPLIFIBER ENGLISH - 29
and press .
5. Choose the appropriate wavelength, 850 or
1300, press
Note
When using a light source other than
SIMPLIFIBER Source, press the to select the
appropriate wavelength. Both units must be
set to the same wavelength.
6. Press, hold and release the button to
set and save the new
reference value. The
reference value will
be displayed.
7. Disconnect the third Launch Cable, located
in the center, without disturbing any connec-
tions to the couplers or adapters.
Measuring Loss using Method C
When Method C was used to set the reference,
proceed as follows:
1. Do not disconnect the launch cables from
the SIMPLIFIBER units.
2. Remove the launch cable located in the