Adjusting Reflected Temperature Compensation Values
As the RTC value changes, the live thermal image will also change as the displayed
temperature changes. The default RTC setting is 100 °C (212 °F)and the adjustment
range is from -50 °C to 600 °C (-58 °F to 1112 °F). RTC is OFF in the factory default
state.
1. From the Home Display, press G (
MENU) three times to navigate to the Adjust
Emissivity and RTC display.
2. Press F (
RTC) to access the Adjust RTC display.
3. Press F (
RTC) to toggle RTC OFF/ON or press H (ADJUST) to access the
Adjust RTC display.
dag121f.bmp