EasyManua.ls Logo

Fluke Ti20 - Specifications

Fluke Ti20
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-1
Appendix D
Specifications
Thermal
Temperature Range............................................ -10 to 350 °C (14 to 662 °F)
Detector Type ..................................................... 80 x 60 thermal element focal plane array (FPA)
Accuracy ............................................................. ± 2 °C or 2 % (whichever is greater)
Repeatability ....................................................... ± 1 % or ± 1 °C (± 2 °F) whichever is greater
NETD (Thermal Sensitivity) ................................ 200 mK
Temperature Indication....................................... 0.1 °C or 0.2 °F
Optical
Field of View (FOV)............................................. Rectangular. 20° Horizontal x 15° Vertical
Minimum Diameter.............................................. 8.1 mm (0.32 in) at 61 cm (24 in)
Optical Resolution (D:S) ..................................... 75:1 or better
Spectral Range ................................................... 7.5 to 14 microns
Target Sighting.................................................... Single Laser (IEC 825/93 Class II, FDA LFR 1040.10 Class II
Instantaneous Field of View................................ 4.4 mrad
Controls
Focus .................................................................. 61 cm (24 in) to infinity
Temperature Scale ............................................. °C or °F selectable
Palettes............................................................... Original, Gray, Rainbow (default), Ironbow, Gray Reverse
Measurement Modes .......................................... Automatic or Manual
LCD Backlight ..................................................... Bright/Dim selectable
Operational
Adjustable Emissivity .......................................... .10 to 1.00 by 0.01
Liquid Crystal Display ......................................... 70.5 mm (2.78 in) x 53.5 mm (2.1 in)
Reflected Background Temperature................... -50 to 905 °C (-58 to 1661 °F)
Ambient Operating Temperature ........................ 0 to 50 °C (32 to 122 °F)
Relative Humidity................................................ 10 to 90 % noncondensing
Storage Temperature.......................................... -25 to 70 °C (-13 to 158 °F) without batteries
Storage Capacity ................................................ 50 images
Electrical
Power.................................................................. Rechargeable battery pack (included)
Battery Life.......................................................... Minimum 3 hours continuous use
Battery Charge Time........................................... 2 hours in the Imager, 1 hour in the external charger (or until the green
LED turns on)
Data Transfer ...................................................... USB interface, transfer time 25 sec for 50 images
Storage Device ................................................... Flash Memory