EasyManua.ls Logo

Fluke Ti20 - C-2. Emissivity Values for Non-Metals

Fluke Ti20
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ti20
Users Manual
C-4
Table C-1. Emissivity Values for Metals (cont)
Emissivity
Material
1.0 µm 1.6 µm 8 to14 µm
Tin (Unoxidized) 0.25 0.1-0.3 n.r.
Titanium
Polished 0.5-0.75 0.3-0.5 n.r.
Oxidized n.r. 0.6-0.8 0.5-0.6
Tungsten n.r. 0.1-0.6 n.r.
Polished 0.35-0.4 0.1-0.3 n.r.
Zinc
Oxidized 0.6 0.15 0.1
Polished 0.5 0.05 n.r.
Table C-2. Emissivity Values for Non-Metals
Material Emissivity
1.0 µm 1.6 µm 8 to 14 µm
Asbestos 0.9 0.9 0.95
Asphalt n.r. 0.95 0.95
Basalt n.r. 0.7 0.7
Carbon
Unoxidized 0.8-0.95 0.8-0.9 0.8-0.9
Graphite 0.8-0.9 0.7-0.9 0.7-0.8
Carborundum n.r. 0.9 0.9
Ceramic 0.4 0.85-0.95 0.95
Clay n.r. 0.85-0.95 0.95
Concrete 0.65 0.9 0.95
Cloth n.r. 0.95 0.95
Glass
Plate n.r. 0.98 0.85
“Gob” n.r. 0.9 n.r.
Gravel n.r. 0.95 0.95
Gypsum n.r. 0.4-0.97 0.8-0.95
Ice n.r. 0.98