21
Using the Toner (TS54 TDR)
Using the SmartTone Function
To use the SmartTone function to identify a wire pair:
1
Turn on the test set, then connect it to a wire pair
in the cable.
2
Make sure that the test set shows 0 VDC.
3
Press M, press D to put the flashing cursor next
to SmartTone, then press M, then press M again
to turn on the toner.
If there is voltage on the line, the display shows
SmartTone Volt Detect Lock Out, and toner will
not operate.
4
At the other end of the cable, use a tone probe to
find the wire pair that has the loudest tone.
5
To make sure you have found the correct pair, put
a short momentarily across the wire pair (Figure 4).
If the tone changes, you have found the correct
pair.
Notes
You cannot hear the tone on the test set.
GOL12.EPS
Figure 4. How to Use the
SmartTone Function (TS54 TDR)
Tone does not change:
Short is not on
the correct pair
Short
Tone changes:
Short is on
the correct pair
Short