EN-5
Note:
Press to select Beep On, Universal Tone On or Beep O.
Press MAIN button to conrm.
When Universal Tone is turned on, the meter guides you
through the blood glucose test using beep tones; it also
outputs the result as a series of beeps.
When Beep is turned o, the alarm function will remain
eective.
During memory deletion, select ”to keep all saved
results.
You may set it up to four reminder alarms.
To turn o the alarm, press or to change On to OFF.
Press MAIN button to conrm.
When the alarm goes o, the device will automatically turn
on. Press to mute the alarm. If you do not press , the
device will beep for 2 minutes then switch o.
If the device is idle for 3 minutes during the setting mode,
it will turn o automatically.
Before Testing
Calibration
You must calibrate the meter every time you begin to use a
new vial of β-ketone/ Total cholesterol/ Uric acid test strips
by setting the meter with the correct code. Test results may
be inaccurate if the code number displayed on the monitor
does not match the number printed on the strip label or strip
foil pack.
How to Code Your Meter (for β-ketone/ Total
cholesterol/ Uric acid test)
1. Insert the code strip when the monitor is o. Wait until the
code number appears on the display.
Note:
Make sure the code numbers on display, code strip, and test
strip vial or foil pack are the same. The code strip should be
within the expiry date; otherwise, an error message may
appear.
2. Remove the code strip, the display will show “OFF”. This
tells you that the meter has nished coding and is ready for
β-ketone/ Total cholesterol/ Uric acid testing.
Checking the Code Number
You need to make sure that the code number displayed on
the meter matches the number on your test strip vial or foil
pack before you proceed. If it matches, you can proceed with