SPECIFICATIONS
System performance
Power source: Four 1.5V AA alkaline batteries
Size of monitor w/o cuff: 147 (L) x 105 mm (W) x 80 mm (H), 500g with
batteries.
Memory: Total 864 memory records
Power saving: Automatic power off if system idle for 3 minutes
(normal mode) or 5 minutes (RF mode).
System operating condition: 10- 40 (50 -104), below 85% RH
Meter storage condition: -20 - 60 (-4 -140), 5 - 95% RH
Power Supply Input:
DC +6V / 1A (max) via Power Plug
Blood glucose measurement performance
Measurement unit: mg/dL or mmol/L
Linear range: 1.1 mmol/L - 33.3 mmol/L
(20 mg/dL - 600 mg/dL)
Precision: ±5 % (CV)
Accuracy: ±15mg/dL when glucose <75mg/dL
±20% when glucose 75mg/dL
Ketone warning: glucose value is over 240 mg/dL
Blood pressure measurement performance
Pressure range: 0 - 300 mmHg
Pulse rate range: 40 -199 beat per minute
Measurement unit: mmHg or KPa
Systolic Measurement Range: 55 mmHg -260 mmHg
Diastolic Measurement Range: 25 mmHg -195 mmHg
Pulse Rate Measurement Range: 40 -199 beats / minute
Maximum inflation pressure: 280 mmHg
Accuracy of Pressure: ±3 mmHg or ±2% of reading
Accuracy of Pulse rate: ±4% of reading
Measurement unit: Either mmHg or KPa
This device has been tested to meet the electrical and safety requirements of:
EN 60601-1, EN 61010-1-2, EN 61010-1, EN 61010-2-101, EN 60601-1-2, EN
61326, EN 300 328