Table 13: Detection of Hypoglycemic and Hyperglycemic Events
Type of Notication Notication Status
15 Minute
Interval
Notication of Hypoglycemic Events
(Low Glucose message)
Detection Rate (%) 77.6
Missed Detection Rate (%) 22.4
False Notication Rate (%) 52.5
Notication of Hyperglycemic Events
(High Glucose message)
Detection Rate (%) 84.7
Missed Detection Rate (%) 15.3
False Notication Rate (%) 10.6
Impending Notication of Hypoglycemic
Events (Glucose Going Low message)
Detection Rate (%) 91.6
Missed Detection Rate (%) 8.4
False Notication Rate (%) 58.8
Impending Notication of Hyperglycemic
Events (Glucose Going High message)
Detection Rate (%) 88.8
Missed Detection Rate (%) 11.2
False Notication Rate (%) 15.5
121