What to know about the Reader’s Built-in Meter:
The FreeStyle Libre 14 day Reader has a built-in blood glucose meter
that is designed to be used only with FreeStyle Precision Neo blood
glucose test strips and MediSense Glucose and Ketone Control
Solution. Using other test strips with the Readers built-in meter will
produce an error or cause the Reader’s built-in meter to not turn on or
start a test. The Readers built-in meter does not have ketone testing
functionality.
The Readers built-in meter is not for use on people who are
dehydrated, hypotensive, in shock, or for individuals in hyperglycemic-
hyperosmolar state, with or without ketosis.
The Readers built-in meter is not for use on neonates, in critically-ill
patients, or for diagnosis or screening of diabetes.
See Using the Readers Built-in meter section for additional important
information on the use of the Reader’s built-in meter.
Where to charge your Reader:
Be sure to select a location for charging that allows the power adapter
to be easily unplugged. Do NOT block access to the charger due to the
potential risk of electrical shock.