Chapter 6: 37
Chapter 6: Service and Repair
6.1 Fuse Replacement
If a fuse is blown for some reason, replace it with a like kind and voltage rating.
Warning: Use of fuses with a different type or rating could result in fire and/or electric
shock hazards.
6.1.1 Line Fuse Replacement
To change the line fuse, first unplug the line cord from the power entry module. Then
using a small screwdriver, pry the top of the module off at the notch near the line cord
socket.
Replace the defective fuse(s) located in the cover holder and replace the cover.
Fuse: (T 630mAL) 0.63A / 250 V~, Type T, IEC 60127-2 sheet III (Slo-Blo Type, Glass,
5mm x 20mm).
(See Section 8 - Rear Panel Safety Markings.)
6.1.2 Connector Board Fuse Replacement
Turn off power to the 8000 and remove the line cord. Remove the wrap-around (see 6.2.1).
Replace the defective fuse(s) located in holder F1 and/or F2 on the connector board.
F1 and F2: (T 315mAL) 315mA / 250 V, Type T, IEC 60127-2 sheet III (Slo-Blo Type, Glass,
5mm x 20mm)
6.1.3 Daughter Board Fuse Replacement
Turn off power to the 8000 and remove the line cord. Remove the wrap-around (see 6.2.1).
Replace the defective fuse(s) located in holder F1 and/or F2 on the daughter board.
F1: (F 3.15AL) 3.15A/250V Type F, IEC 30127-2 sheet II (Fast-Acting Type, Glass, 5mm x
20 mm).
F2: (T 315mAL) 315mA/250V Type T, IEC 60127-2 sheet III (Slo-Blo Type, Glass, 5mm x
20mm)