Chapter 9: 53
Recommended separation distances between
portable and mobile RF communications equipment and the 8000
The 8000 is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the 8000 can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the 8000 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2P
80 MHz to 800 MHz
d = 1.2P
800 MHz to 2.5 GHz
d = 2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
distance d in metres (m) can be estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.