The spindle must already be submerged in the liquid once you confirm the start time.
When the countdown gets to zero, the viscometer will start the calibrating sequence. While the
equipment is calibrating, the following screen will appear (example):
On this screen, the progress bar that appears below the word ¨CALIBRATING¨ displays the status
of this process accompanied by a textual representation of the progress in a percent format.
The exit key MEM/CLEAR’ and allow us to exit to the main menu but never while calibrating (never
while the screen looks like the example just above).
NOTE: Exiting mid-calibration denies the equipment a proper calibration and
therefore it cannot guarantee accurate results.
When the calibration process is over, information on the values of the angles and curvatures of the
calibration are displayed, as it is shown in the following screen:
If the curvature is lower to 2%, hit ‘ENTER’ to confirm the calibration and you will be taken to the
main menu. The User calibration is now stored in the viscometer’s memory.