8.2.4.3 Temperature calibration
Once selected the Factory Calibration or the User Calibration option from the Calibration submenu,
the following screen will appear:
If you select the temperature option (by moving through the menu using the and keys) and
press ‘ENTER’, you’ll be brought to a screen resembling this one:
VERY IMPORTANT: The Test-run should be carried out without a spindle.
Once this message is shown on the screen, we should confirm that the spindle is not connected.
Afterwards, hit ‘ENTER’ and you’ll be brought to a screen resembling this one:
Connect the temperature simulator, using a type B USB connector, to the back of the viscometer
simulating the indicated temperature (in this case 0ºC).
The viscometer’s screen will show the instructions to follow to achieve the calibration of the probe
that measures temperature. You’ll have to connect the PT100 simulator generating an impedance
equivalent to PT100 at 0 degrees Celsius. Once the gauge is connected hit ‘ENTER’ and the
following screen will appear: