EVO EXPERT Manual 27/105
After a few seconds and once the temperature is calibrated to 0 degree Celsius, a second screen of
instructions will appear, containing the following information:
Now, you’ll have to connect the PT100 simulator generating impedance equivalent to a 100ºC
PT100. With the gauge connected and hitting the ‘ENTER’ key, this screen will appear:
After a few seconds, a second screen of instructions will appear, containing the following
information:
Now, you’ll have to connect the PT100 simulator generating impedance equivalent to a 200ºC
PT100. With the gauge connected and hitting the ‘ENTER’ key, this screen will appear: