EVO EXPERT Manual 76/105
Table 11. LCP Adaptor for EVO L
Maximum guideline values in cP (mPa·s)
RPM
LCP
0.01
60000.00
0.3
2000.00
0.5
1200.00
0.6
1000.00
1
600.00
1.5
400.00
2
300.00
2.5
240.00
3
200.00
4
150.00
5
120.00
6
100.00
10
60.00
12
50,00
20
30.00
30
20.00
50
12.00
60
10.00
100
6.00
200
3.00
250
2.40
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of
the selected scale.
Sample Volume = 16 ml.