Interface 1010 Specifications
70
Potentiostatic Mode
Applied Voltage Range
Max
±12
volts
Accuracy
DC zero offset
Gain
Max
1
0.2
mV % setting
mV % setting
Note 18
DC Bias
Typ
±8
V
Scan DAC ranges
Typ
±6.4, ±1.6, ±0.4
V
Drift
Max
<20
µV/C
Note 19
Noise and Ripple
1 Hz to 1 kHz
1 Hz to 200 kHz
Typ
10
20
µV rms
µV rms
Note 20
Galvanostatic Mode
Maximum Full-scale Current
±1000
mA
Minimum Full-scale Current
10
nA
Accuracy
Dominated by current to voltage
converter accuracy shown above
Sig Gen Voltage for Full-scale Current
±3.0
V
Environmental
Operating Temperature Range
0 to 45
C
Relative Humidity
Max
90 (non-condensing)
%
Storage and Shipping Temperature
25 to 75
C
Maximum Shipping Acceleration
30
G
General
Power Input Voltage
Range
46 to 50
V
Power
Max
75
W
Leakage Current
(Floating, Earthed Working Electrode)
Typ
±1
nA
Note 22
Dimensions
24 × 27 × 6
cm
Note 23
Weight
Max
2.0
kg
Note 23
Dimensions of External Power Adapter
Max
13 × 6 × 3
cm
Note 24
Weight of External Power Adapter
Max
300
g
Note 24
NOTES:
1) Measured in galvanostatic mode with a high-power 200 Ω load connected from counter electrode
lead to the working electrode lead. The compliance voltage is measured using an external voltmeter
across the 200 Ω load. Under these conditions, the output current is approximately 115 mA.
2) Measured with a precision 4-terminal 2 Ω load, in potentiostatic mode.