Power LED Blink Codes
82
Error Message This instrument is not authorized for IERange=
The Interface 1010T is able to operate at currents of 100 mA or lower. Higher currents generate this error
message. Decrease the current to continue.
Error Message This instrument is not authorized for ScanRate=
The Interface 1010T is able to scan at rates of 1 V/s or lower. Higher scan rates generate this error message.
Decrease the scan rate to continue.
Error Message This instrument is not authorized for SampleTime=
The Interface 1010T is able to record data with a sample time of 1 ms or higher. Shorter sample times generate
this error message. Increase the sample time to continue.