Frame disassembly
6-15
Document no. 2044677-001
To remove the EMC shield
- remove the screw (T10) and lift the Defib sync unit’s cover
first
- Remove the 14 screws (T10) holding the shield to the frame
unit. Lift the shield carefully up.
To remove the loudspeaker unit
- Remove the two screws (T10) holding the loudspeaker unit to
the frame.
- Disconnect the loudspeaker cable from the CPU board.
To remove the handle guide unit
- Remove the two screws (T10) holding the handle guide unit
to the frame.