Start Up and Operating Control
 
Page 4-12 
Film 
Press the “Film” key to enable film operation. The film LED illuminates when film mode is enabled. In this 
mode, kVp and mAs values can be adjusted manually. Refer to Chapter 7 “Radiographic Film” for details. 
Image Processing Keys 
 Noise Filter 
Press the “Noise Filter” key to select relative noise filter. Noise Filter is only available when real time imaging 
and the image was LIH.  
Recursive  Filter  function  provides  image  noise  reduction.  Higher  levels  of  recursive  filtering  produce 
smoother (or less noisy) image, but cause more lag when imaging moving objects or anatomy.  
Auto/Manual Brightness/Contrast 
Press the “Auto/Manual Brightness/Contrast” key to adjust the brightness and contrast level display.  
The current contrast or brightness level is shown in a numeric value displayed on the bottom right corner of 
the image.   
Auto/Manual Smart Metal 
Press the “Auto/Manual Smart Metal” key to reduce the appearance of metal in the image. It could provide 
optimum image quality even when metal is introduced to the field. The current smart metal level is shown 
in a numeric value displayed on the bottom right corner of the image. When there is no metal in the field, 
please make the smart metal value as 0 to disable it.