Revision C 170 Series Monitor 4-15
2000947-004
Setup Procedures: Customizing the Monitor
Table 4-3. Summary of Service Setup Codes
Code (UA Display)
Setting Or Value
(Primary FHR Display)
Code # Code Description
20 ECG artifact elimination (173, 174 only) 0 = off; 1 = on
21 heartbeat coincidence (172, 173, 174 only) 0 = off; 1 = on
22 fetal movement enable/disable
0 = off
1 = on (if option installed)
23 language
0 = English; 1 = Spanish; 2 = German;
3 = French; 4 = Japanese; 5 = Italian;
6 = Swedish; 7 = Dutch; 8 = Portuguese;
9 = Chinese
24 chart speed 1= 1; 2 = 2; 3 = 3 (cm/min)
25 paper scale 0 = 30–240; 1 = 50–210 (BPM)
30, 40 communications mode (port 1, 2)
0 = HP; 1 = HP w/notes;
2 = ext BP; 3 = factory test; 4 = ext. FSpO
2;
5 = 115 update; 6 = 115 transmit/receive
31, 41 baud rate (port 1, 2)
3
00; 600; 1200; 2400;
48
00; 9600; 19200; 38400
50 remote mark annotation style 0 = FM; 1 = arrow
51 future use (Japanese units only) 0 = off; 1 = on
52 fetal heart rate offset enable/disable 0 = off; 1 = 10 minute auto-revert; 2 = on
53 UA reference default 5, 10, 15, 20, 25 (relative units)
100 recorder alignment 0–255

Table of Contents

Other manuals for GE Corometrics 172

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 172 and is the answer not in the manual?

GE Corometrics 172 Specifications

General IconGeneral
BrandGE
ModelCorometrics 172
CategoryMedical Equipment
LanguageEnglish

Related product manuals