Revision C DINAMAP
®
ProCare Monitor Service Manual A-3
2009381-001
Technical Specifications: Specifications
Specifications
Mechanical
Dimensions Height: 9.7 in (24.7 cm)
Width: 8.6 in (21.9 cm) without Temperature
10.0 in (25.4 cm) with Temperature
Depth: 5.3 in (13.5 cm)
Weight, Including Battery 5.68 lb (2.58 kg)
Mountings Self-supporting on rubber feet
Portability Carried by handle
Classification Information Mode of operation: continuous
Degree of protection against harmful ingress of
water: Drip-proof IPX1
Power Requirements
Power Converter US P/N: 2018859-001
Protection against electrical shock: Class II
AC input: 100-250 VAC/50-60 Hz 12W
DC output voltage: 12VDC at 1.0A
The AC mains power adapter contains a
nonresettable and nonreplaceable fuse.
Monitor Protection against electricalshock: Internally
powered or Class II when powered from specified
external power supply.
DC input voltage: 12 VDC, supplied from a source
conforming to IEC 601-1.
Fuses: The Monitor contains four fuses. The fuses
are auto-resettable and mounted within the
Monitor. The fuses protect the low voltage DC
input, the battery, the remote alarm output, and
the +5 V output on the host port connector.
Battery: 6 volt, 3.3 amp-hours protected by
internal auto-resetting fuse and thermal
protection.
Minimum operation time: 5 hrs (5 min cycle with
adult cuff at 25 °C, SpO
2
active at 60 bpm, Temp in
monitor mode) from full charge.
Time for full
recharge: Approx. 5 hrs from full discharge when
the Monitor is switched off and approx. 8 hrs when
the Monitor is switched on.