Revision C DINAMAP
®
ProCare Monitor Service Manual 6-17
2009381-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
Assembly/Disassembly of FRUs
Monitor Disassembly Procedure
The following procedure is sequential (i.e., you must remove the battery and rear
case to remove the printer, etc.
Battery
1. Remove 4 screws securing the instruction cards.
2. Remove the battery compartment door and adhesive pad.
3. The battery can now be removed.
Rear Case
1. Remove remaining four screws from the bottom of the monitor.
2. Set monitor upright.