Maintenance: Parameter Level Functional Testing
3. Consult the following diagram for pneumatic setup guidelines.
Leakage Testing
NOTE:To enter Service Mode press and hold the Cycle button while pressing
the On/Off button.
1. Turn unit on and enter Service Mode.
2. Press Cycle button and 1 should appear in the min window.
3. Close the valve on the inflation bulb.
4. Using the inflation bulb, inflate the system to 210 mmHg.
5. Allow the system to stabilize for 5 seconds (it is normal to see some
decrease in pressure at this point).
6. Start the stopwatch and record the pressure value.
7. After 60 seconds record the pressure value.
8. The leakage rate is the difference between the first and second readings.
9. Record and verify the leakage rate.
10. Turn the ProCare Monitor off.
MANDREL
2 TUBE
BP CUFF