Maintenance: Cleaning
 Any type of solution that contains ammonium chloride, conductive 
solutions, wax or wax compounds
 Sodium salts
NOTE:Never autoclave or steam clean the monitor, cuffs, or accessories.
Display
To clean the display screen, use a soft, clean cloth dampened with a glass 
cleaner. Never spray the glass cleaner directly onto the display, and never use 
alcohol or petroleum based products. 
Cuffs
General
The cuff must be thoroughly cleaned with the specified detergent before reuse. 
The additional use of household bleach as described below provides at least 
intermediate-level disinfection.
 Apply cuff hose plugs before cleaning.
 The following cleansing procedure was repeated 20 times on DURA-CUF
®
 
Blood Pressure Cuffs and once on SOFT-CUF
®
 Blood Pressure Cuffs without 
affecting the performance of the cuff.
 While this procedure is adequate for cleaning/disinfection, it may not 
remove all stains. 
 Do not immerse hoses.
 Do not immerse cuffs without prior application of cuff hose caps. 
Materials
 Enzymatic detergent such as ENZOL* enzymatic detergent (US) or 
Cidezyme* enzymatic detergent (UK)
 Distilled water
 10% solution of household bleach (5.25% sodium hypochlorite) in distilled 
water
 Soft cloths and soft-bristled brushes
 Spray bottles
Procedure
1. Prepare the enzymatic detergent according to the manufacturer’s 
instructions and the 10% bleach solution, in separate spray bottles.
2. Spray the detergent liberally on device. If the material is dried on, allow the 
cuff to sit for 1 minute. For soil on the soft part of the closure or the cuff 
itself, wipe the material off with a soft cloth. For persistent contamination on 
the soft part of the closure, use a soft-bristled brush to loosen particles. 
Rinse with copious amounts of distilled water. Repeat until no visible 
contamination remains. For soil on the hook part of the closure, use a soft-
bristled brush to remove the material, and rinse with copious amounts of 
distilled water. Repeat until no visible contamination remains.