Electromagnetic Compatibility (EMC): Electromagnetic Compatibility (EMC): ILC-1926
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ILC-1926 is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to assure that the ILC-1926
is used in such an environment.
NOTE:
U
t
is the AC mains voltage prior to application of the test level.
Immunity Test EN 60601 Test Level Compliance Level Electromagnetic Environment – Guidance
Electrostatic
Discharge (ESD)
EN 61000-4-2
± 6 kV contact ± 6 kV contact Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
± 8 kV air ± 8 kV air
Electrical Fast
Transient/Burst
EN 61000-4-4
± 2 kV for power supply
lines
No AC Input No AC Input
±1 kV for input/output
lines
Surge
EN 61000-4-5
± 1 kV differential mode No AC Input Mains power should be that of a typical
commercial or hospital environment.
± 2 kV common mode
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<5% U
t
(>95% dip in
U
t
) for 0.5 cycles
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
EN 61000-4-11
<40% U
t
(>60% dip in U
t
)
for 5 cycles
<40% U
t
(>60% dip
in U
t
) for 5 cycles
Mains power should be that of a typical
commercial or hospital environment. If the
user of the DINAMAP ILC-1926 requires
continued operation during power mains
interruptions, it is recommended that the
Host monitor be powered from an
uninterruptible power supply or a battery.
<70% U
t
(>30% dip in U
t
)
for 25 cycles
<70% U
t
(>30% dip
in U
t
) for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 s
<5% U
t
(>95% dip in
U
t
) for 5 s
Power
Frequency (50/
60 Hz) Magnetic
Field
EN 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristics of a typical location in a
typical commercial or hospital environment.