3-12 DINAMAP
®
ProCare Monitor Service Manual Revision C
2009381-001
Installation: Configuring Your ProCare Monitor
Displayed on Monitor Function
6 (in “min” window)
In “MAP/Cuff” window:
# of remaining saves
possible.
Language
The number displayed in Pulse Rate display window
indicates the language setting. These range from 0 to
20. For example, 0 indicates English. The language
setting is used in printed reports.
0English
1Chinese
2Czech
3Danish
4Dutch
5Finnish
6French
7German
8 Greek
9 Hungarian
10 Italian
11 Japanese
12 Korean
13 Norwegian
14 Polish
15 Portuguese Brazilian
16 Portuguese Continental
17 Russian
18 Slovak
19 Spanish
20 Swedish
Warning: The number displayed in the “MAP/Cuff”
window will decrease each time the “Save Settings”
operation is done. If it reaches zero, a fatal 975 alarm
(calibration space exhausted) results and the unit will
have to be serviced before operation will be possible