rev.1.1 Page 11-23
Cleaning Procedure 1 2 3 4 5 6 7 8
PROBLEM
C, E chamber
metallic scaling
Yes
D, C, E chamber
metallic scaling
Yes
Organic fouling Yes
Organic fouling and
metallic scaling
Yes
Biological fouling Yes
Biological fouling
and metallic scaling
Yes
Heavy biological
fouling
Yes
Heavy biological
fouling and metallic
scaling
Yes
Severe biological
fouling
Yes
Severe biological
fouling and metallic
scaling
Yes
The table below summarizes the main steps that are performed for each procedure. These
steps are provided in detail in subsections which follow.
Cleaning
Procedure
1 2 3 4 5 6 7 8
Instruction 1 C Acid Acid Caustic Acid Brine Acid Brine Acid
Instruction 2 Rinse Rinse Rinse Caustic Rinse Brine Rinse Brine
Instruction 3
- Regen Regen Rinse Santize
A or B
Rinse Santize
A or B
Rinse
Instruction 4
- - - Regen Brine Santize
A or B
Caustic Santize
A or B
Instruction 5 - - - - Rinse Brine Rinse Caustic
Instruction 6 - - - - Regen Rinse Regen Rinse
Instruction 7 - - - - - Regen - Regen
Notes
:
1) The cleaning procedures are written for an E-Cell System
containing a single Stack. For
larger systems scale up proportionally. Maintain chemical concentrations by multiplying
all water and chemical amounts by the number of Stacks.
2) E-Cell System™ cleaning requires up to 8 hours per cleaning procedure.
3) After cleaning, the E-Cell Stacks™ must be regenerated (except Procedure 1).
Regeneration requires up to 16 hours.