4400-0024-EN Rev A
2.  Leakage current at the AC power cord 
Equipment 
Safety analyzer with test lead 
enFlow Controller
 
Purpose 
This test is run to check the chassis leakage in microamps. For the purpose of this test, the Controller is the chassis. 
Procedure 
1.  Plug the power cord of the Controller into the safety analyzer. 
2.  Turn the knob on the safety analyzer to the chassis leakage function. 
3.  Power the safety analyzer by plugging in and setting the MAINS power to ON. 
4.  Record the polarity and ground readings for the Controller for both power on and o scenarios for the following congurations:
Allowable values of continuous LEAKAGE CURRENTS, in μA.
ENCLOSURE LEAKAGE CURRENT TYPE BF
Normal polarity — normal ground  Normal Condition  100
Reverse polarity — normal ground  Single Fault Condition  500
Reverse polarity — open ground  Double Fault Condition  500
Normal polarity — open ground  Single Fault Condition  500
3.  Leakage current of the Warmer to the saline in the IV line 
Equipment 
enFlow Warmer and Disposable Cartridge 
Safety analyzer with ECG leads 
Saline 
Wire 
Extension set with a non-venting cap
Fill with saline
Wire