The green indicator shows:
•
when flashing — power is on
•
when continuously lit — power is on and connection with UNICORN is established.
When lit, the yellow indicator shows that Fraction Collector Frac-950 is running.
When connected to UNICORN, several beeps are heard after which Fraction Collector
Frac-950 starts to perform an automatic calibration process. During this process, the
tube holder moves rapidly in several directions. When calibration is completed, the tube
holder stops in the home position.
The safety bar must be folded up for Fraction Collector Frac-950 to start up.
Note:
Operating Fraction Collector
Frac-950
Fraction Collector Frac-950 is controlled from a PC running UNICORN version 3.21 or
higher for standard mode or version 4.0 or higher for prep mode. For ÄKTAmicro, UNICORN
version 5.0 or higher is recommended. Control of Fraction Collector Frac-950 can be
achieved automatically from a method, or manually via the functions available in
UNICORN.
Using Fraction Collector Frac-950 in a method is described in the ÄKTA design Optional
Configuration User Manual.
The following functions are available for operating Fraction Collector Frac-950 from
UNICORN:
MethodManual
FractionationMan_Fractionation
Fractionation_StopFractionation_Stop
Feed_TubeFeed_Tube
Peak_FractionationMan_Peak_Fractionation
Peak_Frac_StopPeak_Frac_Stop
AccumulatorWashAccumulatorWash
-Ignore_LastTube
Reset_Frac_PositionReset_Frac_Position
(PeakFrac Parameters)(PeakFrac Parameters
40 Frac-920 and Frac-950 Operating Instructions 28-9622-09 AC
4 Operation
4.3 Fraction Collector Frac-950