Chapter 2: Operation
Page 32 HygroPro User’s Manual
2.3.2 Entering Sensor Calibration Data (cont.)
Select the desired units and press enter.
Use the arrow keys to enter a value and
press enter.
Repeat the previous steps until all of your Hygro Curve data points
have been entered.
To save the data you have entered, scroll
to the bottom of the Hygro Curve
options list to the SAVE HYGRO TABLE?
section. Then, select Save and press
enter.
IMPORTANT: Any data not saved at the above prompt will be lost.
Repeat the above process to enter any available Press Curve and
CS Table data points.
Note: The CS Table is required only if ppm
w
measurements will be
made. Consult GE for the table values to use for your
application.