EasyManua.ls Logo

GE HygroPro - 4.2 Electrical

GE HygroPro
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4: Specifications
Page 44 HygroPro User’s Manual
4.2 Electrical
Power
Input: 12 to 30 VDC (loop-powered, customer supplied)
Output: 4 to 20 mA analog, RS485 digital
Output Resolution: 0.01 mA/12 bits
Maximum Load Resistance: = (PSV x 33.33) – 300,
where PSV = Power Supply Voltage
Example: Given a 24 VDC Power Supply,
Max. Load Resistance  33.33
) – 300 = 500
Cable: 6 ft (2 m), standard (consult GE for custom lengths)
Input Parameters for Loop-Powered Intrinsic Safety
U
i
= 28V P
i
= 0.653W L
i
= 62µH
I
i
= 93.3mA C
i
= 0
RS485
U
i
= 3.72V P
i
= 212mW L
i
= 62µH
I
i
= 228mA C
i
= 67µF at 5.36V

Related product manuals