HygroPro User’s Manual Page 37
Chapter 3: Service & Maintenance
3.3.1 Preparing to Clean the Probe (cont.)
To clean the moisture probe, the following items are required:
Two glass (NOT metal) containers with approximately 300 ml of
reagent-grade hexane or toluene in each container.
One glass (NOT metal) container with approximately 300 ml of
distilled (
NOT deionized) water.
IMPORTANT: Make sure the containers are deep enough to completely
submerge the moisture probe. Do not place the
transmitter module into the solvents. Insert only the
moisture sensor into the solvents.
Rubber or latex gloves
An oven set at 50°C ± 2°C (122°F ± 3.6°F)
1-1/8” wrench
3.3.2 Replacing the RTE
To maximize HygroPro performance, GE recommends recalibration
of the aluminum oxide moisture sensor on the Replaceable
Transducer Element (
RTE) every 6 to 12 months. The optimum
interval depends on the specific application. To accomplish this,
either return the RTE to GE for recalibration or install a new RTE.
The HygroPro electronics will automatically read and store the
calibration data whenever a new or recalibrated RTE is installed.
IMPORTANT: The programmed probe calibration data should not be
changed without consulting GE.