7 - 14 Section 7-3 - Gathering Troubleshooting Data (R1 to R4)
7-3-7 Capturing Network Logs with Network Sniffer
Logs collected should be stored under d:/log/Sniffer folder to ensure they form part of general log
collection Alt+D or Collect Log.
1.) Press Alt+N.
2.) On the network sniffer screen, select Capture -> Options.
Figure 7-13 Network Sniffer Screen
Select the interface that will be receiving packets (Network card A).
NOTE: Remember, if the device has a DVR, it will show up in the list. Select either Intel Network
connection or the Wireless network adapter.