DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-3 - Gathering Troubleshooting Data (R1 to R4) 7 - 15
7-3-7 Capturing Network Logs with Network Sniffer (cont’d)
Use filters to limit the captured data. Filter by IP or by IP and port number.
By IP only: press on Capture Filter.
3.) Select IP only.
Use the IP address of the system or the one from the DICOM device under test, on the Filter string field,
using the following syntax: (example) host 3.62.12.33 as shown in below, press OK.
Figure 7-14 Capture Filter IP Only