LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-115
8-4-1 Installing Base Image Software (cont’d)
10.)Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software Disk
from the CD/DVD ROM drive.
11.)After removing the disk from the drive, Press ALT + CTRL + DEL key.
NOTE: After Installing Base Image Software, the “General failure reading drive A. Abort, Retry, Fail?”
message can be appeared, but Ignore it! All procedure would be accessibly done.
12.)The following message box can be displayed after the SYSTEM boot up. Press Yes.
Figure 8-170 Installation Completion
WARNINGWARNING
Remove the disk from the drive. Otherwise you will be encountered the Base Image
Software installation screen iteratively.
Figure 8-171 System Setting Change