GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-19
3-6-6 Connecting the USB Flash Memory Stick
The USB Flash Memory Stick may be connected to an accessible USB port of the LOGIQ™ S7 Expert/
Pro system. Refer to Section 3-9-2 "External I/O Connectors" .
An external USB Flash Memory Stick can be connected once the system is powered ON, or after
shutdown. The LOGIQ™ S7 Expert/Pro, Windows detects the device and automatically installs a driver.
During this process several dialogs may pop up, starting with the „Found New Hardware” dialog.
NOTE: Memory drives or sticks may be sensitive to EMC interference.
This may affect system performance and/or image quality.
3-6-7 Connecting the external USB Hard disk (Handydrive)
The external “Handydrive” HDD may be connected to an accessible USB port of the LOGIQ™ S7
Expert/Pro system. Refer to Section 3-9-2 "External I/O Connectors" .
An external USB Hard Disk Drive can be connected once the system is powered ON, or after shutdown.
The LOGIQ™ S7 Expert/Pro, Windows detects the device and automatically installs a driver. During
this process several dialogs may pop up, starting with the „Found New Hardware“ dialog.
NOTE: Memory drives, sticks or HDD drives may be sensitive to EMC interference.
This may affect system performance and/or image quality.
!! NOTICE:
Before disconnecting an external USB-device (e.g., USB Stick), the system has to be informed about
the removal of the device! For this purpose press the EJECT
button on the keyboard.
For further details refer to: Section 3-6-8-2 "External USB-Devices - Disconnection" on page 3-20.
!! NOTICE:
Before disconnecting an external USB-device (e.g., USB Hard disk), the system has to be informed
about the removal of the device! For this purpose press the EJECT
button on the keyboard.
For further details refer to: Section 3-6-8-2 "External USB-Devices - Disconnection" on page 3-20.