Diagnostics/Troubleshooting
8-22 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Assessment Utility (continued)
8. In the Diagnostic-> Service Diagnostics-> Assessment
Utility-> Probe Assessment to check the probe.
If there is no broken element and the probe is acceptable for
diagnostic, the following result window displays.
Figure 8-13. Probe Assessment Pass