Diagnostics/Troubleshooting
8-24 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Probe Pass/Fail Criteria
Pass
• No dead element
• Dead elements on the edge (#0, #1 or #end-1, #end)
Fail
• Max 2 weak elements on the scan line:
-3dB<elements<-2dB
• Any one of the elements in one scan line: element<-3dB