Maintenance task schedule
LOGIQ V2/LOGIQ V1 – Basic Service Manual 10-7
5610739-100 English Rev.10
NOTE: The maintenance may require specialized equipment to
complete.
NOTE: The periodic maintenances are not mandatory. The table above
is for reference only.
Clean LCD
Console Leakage Current
Checks
See Notes Twice Annually
Peripheral Leakage
Current Checks
See Notes Twice Annually
Surface Probe Leakage
Current Checks
See Notes Twice Annually
Endocavity Probe Leakage
Current Checks
See Notes Quarterly Annually
Surgical Probe Leakage
Current Checks
See Notes Quarterly Annually
Measurement
Accuracy Checks
See Notes Twice Annually
Functional Checks See Notes also after corrective
maintenance
Table 10-1: Customer Care Schedule (Continued)
Service at Indicated
Time Daily Weekly Monthly
Per
Facilities
QA
Program