EasyManua.ls Logo

GE LOGIQ V5 Expert

GE LOGIQ V5 Expert
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Global Service User Interface (GSUI)
LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-15
5726264-100 English Rev.8
7-4-2 Control Frame
Contains the user interface elements used for:
• Diagnostic control, and
• Operator feedback
7-4-2-1 Button
This button has two modes each with appropriate text:
Execute - to start the diagnostic, and
Abort - to stop a diagnostic
The button can also be disabled.
7-4-2-2 Loop Count
This is an editable text field that will only accept numeric values
with 4 digits or less. When the button is configured as an
“execute” button and pressed, the loop count field will be
queried to determine the number of times to execute the
diagnostic.
7-4-2-3 Progress Indicator
Displays a graphical progress indication to the user.
7-4-2-4 Short Text Message
Displays either a starting message or aborting message, as well
as the diagnostic completion status.
7-4-2-5 Background Color
Initially gray, the Control Frame background color changes upon
completion of a diagnostic to indicate completion status.
• Fail = Red
Pass = Green
Neither pass nor fail = Set back to Gray (for example, final
code status is Aborted).

Table of Contents

Related product manuals