Diagnostics/Troubleshooting
7-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
7-4-3-2-16 Probe Pass/Fail Criteria
Pass
No dead element
Dead elements on the edge (#0, #1 or #end-1, #end)
Fail
Max 2 weak elements on the scan line:
-3dB<elements<-2dB
Any one of the elements in one scan line: element<-3dB
7-4-3-3 Restart the system after diagnostics
Always shutdown the system and reboot after a diagnostics
session.