EasyManuals Logo

GE LOGIQ V5 Expert User Manual

GE LOGIQ V5 Expert
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #216 background imageLoading...
Page #216 background image
Diagnostics/Troubleshooting
7-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
7-4-3-2-16 Probe Pass/Fail Criteria
Pass
• No dead element
• Dead elements on the edge (#0, #1 or #end-1, #end)
Fail
• Max 2 weak elements on the scan line:
-3dB<elements<-2dB
• Any one of the elements in one scan line: element<-3dB
7-4-3-3 Restart the system after diagnostics
Always shutdown the system and reboot after a diagnostics
session.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 Expert and is the answer not in the manual?

GE LOGIQ V5 Expert Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5 Expert
CategoryMedical Equipment
LanguageEnglish

Related product manuals