EasyManua.ls Logo

GE LOGIQ V5 Expert

GE LOGIQ V5 Expert
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Loading the software
LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-27
5726264-100 English Rev.8
8-3-8-3 Option Strings Check
NOTE: After the system software loading completion, please check the
option strings to ensure that the options are activated and
working.
1. Reboot the system.
2. Press Utility on the control panel.
3. Select Admin -> System Admin.
4. Ensure that all the installed option keys are displayed and
the status of Options are valid.
The status “Valid” means the option keys are activated
and working.
The status “disabled” means the option keys are not
activated and not working. Check if the option is
installed and if the serial number and option key are
correct.
Figure 8-24. Software Option

Table of Contents

Related product manuals