GE
Medical
Systems
MAC
1200
V
1.2
Page
34
Information
Technologies
227
492
21
Rev
F
3.6
Test
results
Pressing
the
“4"
key
triggers
the
output
of
the
test
results.
The
output
of
the
test
protocol
supplies
the
following
information:
m
Time
and
date
m
Device
serial
number
m
Device
model
m
Software
release
m
ECG
time
constant
m
Battery
voltage
E
test
results
m
Memory
used
(ROM,
RAM)
m
base
address
E
size
Following
output,
the
initial
display
menu
reappears
automatically.
The
identification
of
the
Hardware
Version
is
given
by
the
item
“Remote
Control
yes/no”.
“No”
indicates
Hardware
Version
V
1.0
(no
remote
control),
“yes”
indicates
Hardware
Version
V
1.1
or
V1.2
This
message
is
intended
only
for
the
use
of
the
individual
or
entity
to
which
it
is
addressed.
As
it
contains
confidential
information,
any
use
distribution
or
copying
is
strictly
prohibited
if
you
are
not the
addressee.
3.7
Self-test
The
test
results
elucidated
in
Section
3.6
are
ascertained
during
the
self-test,
which
is
always
performed
on
power-up.
Should
errors
be
detected,
a
message
appears
on
the
display
after
the
test,
indicating
the
possible
errors.
The
following
error
codes
are
used
to
identify
the
error.
Error
codes
The
following
error
codes
appear
on
the
display
together
with
the
message
“Self-test
failed”:
ERROR_CODES:
0
-
reserved
1
VEKT
-
error
in
vector
table
2
DRAM
-
DRAM
error
3
SRAM
-
SRAM
error
4
-reserved

Table of Contents