GE
Medical
Systems
MAC
1200
V
1.2
Page
44
Information
Technologies
227 492
21
Rev
F
Replacing
the
primary
fuses
in
the
mains
input
module
Warning:
Only
use
fuses
indicated
on
the rating
label!
(see
Section
8,
Paris
List)
Replacing
the
battery
Disconnect
and
lift
out the
battery
compartment
on
the
underside
of
the
device
by
raising
the
middle
fastening
catch.
Remove
the
battery
and
disconnect
the
battery
plug
BATT_IN/
from
the
PCB
Battery
Charge.
Warning:
Only
original
GE
Medical
Systems
Information
Technologies
batteries
may
be
used!
(see
Section
8,
Parts
List)
Defective
NC-batteries
or
empty
batteries
should
be
disposed
of
in
accordance
with
the
applicable
legal
stipulations
or
returned
to
the
factory.
Batteries
returned
to
the
factory
should
be
labeled
“For
disposal”.
Before
putting
in
the
new
battery,
insert
plug
BATT_IN/.
Push
in
battery
compartment,
secure
by
pressing
on
the
fastening
catch.
Replacing
the
PCB
Battery
Charge
After
opening
the
device,
first
of
all
open
the
battery
compartment
and
disconnect
the
battery
connector
BATT_IN/
(see
"Replacing
the
battery’).
Then
disconnect
the
connector
BATT/
from
the
PCB
Control.
Undo
the
fastening
screw
next
to
the
inductor
L1
on
PCB
Battery
Charge.
PCB
Battery
Charge
has
one
soldered
fuse:
Si1:
Battery
fuse
No
adjustments
on
the
PCB
Battery
Charge
are
necessary.
Replacing
the
PCB
Control
Before
replacing
the
PCB,
if
still
possible,
printout
the
settings
configured
by the
user
and
check
the
options
enabled.
After
opening
the
device
first
of
all
disconnect
the
plug-in
connector
POSUP/
to
the
power
supply
unit,
and
the
connector
BATT_IN/
from
the
PCB
Battery
Charge.
Make
sure
that
the
connector
BATT_IN/
makes
no
contact
with
the
coated
shell,
while
disconnected.
Then