RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
PERIODIC MAINTENANCE
10–5
REV 24
10–2 PERIODIC MAINTENANCE PROCEDURE (Continued)
Note
The CRT Filter Clamps and Escutcheon mounted on the monitor for the RTfino/RT3200 Advantage–
II/RT–x200 console are changed to strengthen the mounting grip of the CRT filter on the escutcheon.
With these changes, the CRT Filter Clamp Kit are provided to improve the installed base systems.
This is:
PART NAME QTY PART # REMARKS
CRT Filter Clamp P Kit 1 2152056 For CRT FILTER (plastic
filter, P9514BY)
CRT Filter Clamp G Kit 1 2152119 For VDT FILTER (glass
filter, P9500SF)
8. Perform electrical safety tests on the following components:
a. Hospital outlet (Facility power wiring test).
b. All transducers (Transducer source leakage current test).
c. Console and peripherals (Chassis source leakage current test and ground continuity test).
See Chapter 3 of the RTfino/RT3200 Advantage–II/RT–x200 Service Manual for test procedures.