RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–24
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
PUSH BRACKET
MONITOR ARM
CLAMP PLATE
STUD
SCREW
FIXING SCREWS
BOTTOM VIEW – MONITOR ARM
ILLUSTRATION 3–14
10. Remove the bellows both from the monitor arm and the main unit. Some force may have to be applied to loosen
the adhesive holding the bellows.
11. Unscrew the four Allen screws to remove the monitor arm from the pipe.
MONITOR ARM
ALLEN SCREWS
PIPE
BELLOWS
RAISED MONITOR ARM
ILLUSTRATION 3–15
12. Unscrew the screws from the front cover, front plate, front panel, rear cover, and remove them so that the inside of
the main body will be exposed.