RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–34
REV 21
SECTION (A) Data Recording (continued)
2–9. Record Probe Data.
The number of connectable probes differs depending on the Version (Style).
The parameter setting image in Illustration 3–24 may not be the same in every case. Illustration 3–24 is
shown as Version 2(Style B) . However the procedure is the same as the following:
i. Press the <Control>, S , ! and then the <Enter> key and verify that the CRT displays “SERVICE !” in
upper–left corner of the screen.
ii Press the S , E and <Enter> keys.
iii All the probe parameter settings appear on the screen. If they are different from those in the attached
list, record the screen (or write down any different items directly onto illustration 3–24).
iv Press the <Control> key.
PRB G DR FOCUS SCL DIR INV MAP HFR
MZ 50 48 3 1.5 RVS NOR NOR OFF
RA 50 54 3 1.0 NOR NOR NOR OFF
L 50 54 3 1.0 NOR NOR NOR OFF
S 50 54 3 1.0 NOR NOR NOR OFF
K 50 54 3 1.0 NOR NOR NOR OFF
A 50 54 3 1.0 NOR NOR NOR OFF
B 50 54 3 1.0 NOR NOR NOR OFF
C 50 54 3 1.0 NOR NOR NOR OFF
D 50 54 3 1.0 NOR NOR NOR OFF
E 50 54 3 1.0 NOR NOR NOR OFF
F 50 54 3 1.0 NOR NOR NOR OFF
G 50 54 3 1.0 NOR NOR NOR OFF
H 50 54 3 1.0 NOR NOR NOR OFF
I 50 54 3 1.0 NOR NOR NOR OFF
J 50 54 3 1.0 NOR NOR NOR OFF
R 50 54 3 1.0 NOR NOR NOR OFF
CV 50 54 3 1.0 NOR NOR NOR OFF
CA 50 54 3 1.0 NOR NOR NOR OFF
CB 50 54 3 1.0 NOR NOR NOR OFF
CC 50 54 3 1.0 NOR NOR NOR OFF
G50D54T60
CONTROL : SE
GE
SERVICE!
13/12/91
13:55:26
CA 5MHz
*1.0
MANUFACTURE’S PRESET PARAMETER
ILLUSTRATION 3–24