RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–22
REV 21
3–8 REPLACEMENT
3–8–1 GAS SPRING REPLACEMENT
1. Raise the monitor to the highest position.
2. Unscrew and then remove the bottom (sub) cover. (See Illustration 3–10)
BOTTOM (SUB)
COVER
SCREW
ASSEMBLED MONITOR BOTTOM VIEW
ILLUSTRATION 3–10
3. Pull out the monitor cable and disconnect the connector. (See Illustration 3–11)
4. Remove two rubber plugs and unscrew the two neck pins and then lower the monitor to the lowest position. (See
Illustration 3–12)
5. Turn the monitor until the groove in the rotation slider (See Illustration 3–11) comes to the position of the pin hole.
(See Illustration 3–12)
6. Pull the monitor out of the monitor arm. Ensure that the nylon washer remains in place in the monitor arm. (See
Illustration 3–12)
7. Unscrew the three screws fixing the monitor arm cover under the monitor arm and remove monitor arm cover.
(See Illustration 3–13)