RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–52
REV 21
SECTION (E) (continued)
11–9.Probes parameter settings.
i. Connect one of the probes used by the doctor and modify the value of the probe parameter settings in
order to obtain the correct values (Gain, Dynamic, Focus, Scale, Direction, Inverse, Map and HFR).
Refer to the Operator Manual to preset the Probes parameter.
ii Press the <Control>, B and then the <Enter> key. Now, all the settings for one probe are completed.
iii Repeat the same setting procedure for the other probes.
iv Press the <Control> key.
11–10.Additional anatomical features settings.
i. Enter the Additional anatomical features recorded in step 2–10.
Refer to the Operator Manual to enter Additional anatomical features.
11–11.User Programmable Comments Library.
i Enter code and text recorded in step 2–11.
Refer to the Operator Manual to Enter code and text.
12. Dispose of excess material.
13. This completes the procedures.