Page #107 background image
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Viewer
11-viewer.fm Page no. 107 Chapter 11
3-1 Annotation Content
Annotations available on the image are:
- Patient information (name, ID, date of birth, age, sex)
- Acquisition date and time
- Hospital information
- X Ray parameters (Entrance Skin Exposure, Average Glandular Dose, mAs, kV, Focal Spot, Track,
Filter, Thickness, Compression force, exposure mode)
- Anatomical information (Laterality, View name, magnification factor)
- Processing information (type*, date and time)
- Display parameters (window width, window level, zoom factor, image size, FOV)
- Measurements (annotations associated with measurements)
* Processing type depends on settings made in Medical Application preferences panel (see section 8-3
Medical Application preferences on page 75 in Chapter 8 Browser). In case Premium View is not
enabled, Thickness Equalization applies, or Auto-contrast for specific cases.
Cases when Thickness Equalization cannot apply are described in section 12 Processing on page 135 in
Chapter 12 Image Acquisition Procedure.
Processing type displayed on the image can be one of the following:
•PROC_0, meaning that:
- FineView processing not applied before the raw image
- Auto-Contrast processing applied on the raw image
•PROC_1, meaning that:
- FineView processing not applied before the raw image
- Thickness Equalization processing applied on the raw image
PREMIUM_VIEW, meaning that:
- FineView processing not applied before the raw image
- Premium View processing applied on the raw image
PROC_0_FV, meaning that:
- FineView processing applied before the raw image
- Auto-Contrast processing applied on the raw image
PROC_1_FV, meaning that:
- FineView processing applied before the raw image
- Thickness Equalization processing applied on the raw image
PREMIUM_VIEW_FV, meaning that:
- FineView processing applied before the raw image
- Premium View processing applied on the raw image
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals